All instruments are autoclaved at up to 134 °, can be sterilized in the hot-air between 180 ° and 220 °.
Instruments are to be handled according to its method of sterilization. Pay attention to the operating instructions of the device. All instruments are both sterilized with chemical, or thermodynamic process. Hot air, autoclave, etc.
• New instruments must be degreased before first use by cleaning
• Disinfect Used instruments first, then clean, rinse, dry and sterilize
• Hinges should be well cleaned and treated with a conditioning spray after drying
• sort out defective instruments prior to sterilization, as well as in stainless steel rust can be transferred
• Do not store instruments in alcohol or disinfectant solution
Avoid base metals, which can spill over to the sterilizer instruments. e.g. cleaning with wire brushes, or common sterilization with chrome metals.
The storage can be made in sterile or non-sterile, for example, in trays. It should be noted moisture as far as possible be avoided. Coarse dirt is completely removed. Most instruments with plastic attachments are easy to sterilize. A review of the turnover is recommended every 50 cycles. Instruments with tungsten carbide inserts have a considerably longer life time. A gentle and dedicated treatment of the instruments, prolongs life. Once instruments are suitable special sterilization methods are usually applied for not Sterilisationsmethoden.Im field of transplantation surgery. Ask if in doubt with us after. The Garantieranspruch expires if not used properly.
FASA GROUP also offers you a cost-effective and very fast grinding and repair service (also for third-party tools). The return takes place within a few days!
How to care for implants
For metallurgical, mechanical and design reasons manufacturers, as well as from various materials should never be combined. Materials used are stated in the catalog and on the product labels.
CAUTION: Repeated bending back and forth leads to fatigue or fracture of the implant. Nicks and bruises reduce the mechanical strength also greatly!
Notes and Warnings:
Implants may not be re-used! Explanted implants must be supplied to the hospital's disposal werden.Um full traceability to ensure that products and batch number of the implant used must be attached to the OP report.
NOTE: In the application of partial and Vollgewindebohrdrähten with small diameters, they can break easily if used improperly. We are unable to accept any liability.
Examination upon receipt and before use
Check the implants after receipt for shipping damage and to condensate in the shipping container and remove the implants of the overpack. Examine the implants prior to their use in the operating room for discoloration, cracks, nicks, and other damage that could be caused by improper storage and sterilization.
Advice on storage compatibility
Store the implants clean and dry at 15 - 35 ° C. Be especially careful that no chemicals are in close proximity.
Cleaning and sterilization
Before applying the implant, the original packaging must be removed and a complete treatment response (cleaning, sterilization) are carried out by qualified personnel. The implants can be steam sterilized according to ISO 17665 (temperature 134 ° C, holding time: ≥ 5 minutes at a pressure of 3 bar).
The instructions of the sterilizer manufacturer must be observed!